Research on Property of MCP-PMT's Pulse Output Linearity
doi: 10.11804/NuclPhysRev.31.02.177
- Received Date: 1900-01-01
- Rev Recd Date: 1900-01-01
- Publish Date: 2014-06-20
-
Key words:
- MCP-PMT /
- pulse linear current /
- double pulse method
Abstract: The double pulse method was applied to study the pulse linear current property of R3809U-52 type MCP-PMT (micro-channel plate photo-multiplier tube) from amamatsu Photonics Co.Ltd. When the FWHM of pulse are 20 and 100 ns, the pulse linear currents are 150 and 25 mA, respectively. According to the experiments results, we deduced the functions of pluses' FWHM lower than 100 ns with linear current, which extends the range of parameters available.
Citation: | CHEN Yuyu. Research on Property of MCP-PMT's Pulse Output Linearity[J]. Nuclear Physics Review, 2014, 31(2): 177-182. doi: 10.11804/NuclPhysRev.31.02.177 |