Study of Double-Layer Multi-Wire Proportional Chamber based on LC Delay Circuit
-
ZHANG Xing,
-
HU Rongjiang,
-
LU Chengui,
-
YANG Herun,
-
MA Peng,
-
GUAN Yuanfan,
-
WEI Xianglun,
-
DUAN Limin,
-
WANG Jiansong,
-
ZHENG Yong,
-
Graphical Abstract
-
Abstract
A double-layer Multi-Wire Proportional Chamber (MWPC) with 100 mm×80 mm active area has been developed for Radioactive Ion Beam Line in Lanzhou(RIBLL Ⅱ). The position information has been obtained by a LC delay circuit readout method. Being called of double-layer MWPC, it is composed of two full MWPCs, each consists of an anode plane sandwiched between two parallel cathode planes, anode wires and cathode wires are perpendicular to each other. In order to improve the detection efficiency, the cathode wires of corresponding position are combined to enhance inductive signal. The cathode wires of two full MWPCs are orthogonally placed to give two coordinates of the incident particles. The sensitivity uniformity is also found to be relatively good and the position resolution of X, Y direction are measured to be 199.9 and 154.0 μm using a 55Fe-5.9 keV X ray, respectively. The position linearity of X, Y direction are as good as 0.999 for whole sensitive area of the detector. The detector has been used successfully on the experiment at RIBLL Ⅱ.
-
-