Calibration of Neutron Detection Efficiency of Li-glass Detector

doi: 10.11804/NuclPhysRev.30.02.156
  • Received Date: 1900-01-01
  • Rev Recd Date: 1900-01-01
  • Publish Date: 2013-06-20
  • In order to accurately measure the neutron capture cross section in the energy range of keVMeV, 4 πgamma-ray total absorption facility (GTAF) is being constructed at China Institute of Atomic Energy (CIAE). The lithium glass detector will be used as a neutron beam monitor for GTAF. The detection efficiency of the lithium glass detector at two incident neutron energy points (250, 565 keV) was calibrated in 5SDH-2 accelerator, and the relative detection
    efficiency was simulated by GEANT4 and MCNP code. By the normalization of the experimental data and simulation result, the neutron detection efficiency curve of the lithium glass detector between 10 keV and 1 MeV was obtained. This work will be important to convert the Time-of-flight spectrum that be measured by Li-glass detector to the energy spectrum of neutron beam, and provide the new method for calibration of detection efficiency.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(2277) PDF downloads(573) Cited by()

Proportional views

Calibration of Neutron Detection Efficiency of Li-glass Detector

doi: 10.11804/NuclPhysRev.30.02.156

Abstract: In order to accurately measure the neutron capture cross section in the energy range of keVMeV, 4 πgamma-ray total absorption facility (GTAF) is being constructed at China Institute of Atomic Energy (CIAE). The lithium glass detector will be used as a neutron beam monitor for GTAF. The detection efficiency of the lithium glass detector at two incident neutron energy points (250, 565 keV) was calibrated in 5SDH-2 accelerator, and the relative detection
efficiency was simulated by GEANT4 and MCNP code. By the normalization of the experimental data and simulation result, the neutron detection efficiency curve of the lithium glass detector between 10 keV and 1 MeV was obtained. This work will be important to convert the Time-of-flight spectrum that be measured by Li-glass detector to the energy spectrum of neutron beam, and provide the new method for calibration of detection efficiency.