Advanced Search

LAN Hai-yang, ZHENG Huaqing, LI Gui, #, . Improved Corrected Algorithm of Wedge Factors in Accuracy Radiotherapy[J]. Nuclear Physics Review, 2010, 27(1): 77-81. doi: 10.11804/NuclPhysRev.27.01.077
Citation: LAN Hai-yang, ZHENG Huaqing, LI Gui, #, . Improved Corrected Algorithm of Wedge Factors in Accuracy Radiotherapy[J]. Nuclear Physics Review, 2010, 27(1): 77-81. doi: 10.11804/NuclPhysRev.27.01.077

Improved Corrected Algorithm of Wedge Factors in Accuracy Radiotherapy

doi: 10.11804/NuclPhysRev.27.01.077
  • Received Date: 1900-01-01
  • Rev Recd Date: 1900-01-01
  • Publish Date: 2010-03-20
  • In accuracy radiotherapy, it is difficult to calculate dose accurately during wedge fields. In this paper, an improved corrected algorithm of wedges based on the lineattenuation was proposed. The lineattenuation coefficient of wedges was calculated firstly, then the actual path of radial beam in the wedge was calculated by the “linear fitting” according to the sizes of wedges section. Finally the wedge factors could be determined accurately, and the dose could be corrected. The data of the 45 degree wedge in the report of AAPM 55 were used to make the calculation and verificaton. Using this algorithm,amount of measured data were decreased, the accuracy of calculation could be assured, and the job of physician could be reduced,so this algorithm could be applied in the treatment planning system as an effective calculation method of wedges field.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(2235) PDF downloads(811) Cited by()

Proportional views

Improved Corrected Algorithm of Wedge Factors in Accuracy Radiotherapy

doi: 10.11804/NuclPhysRev.27.01.077

Abstract: In accuracy radiotherapy, it is difficult to calculate dose accurately during wedge fields. In this paper, an improved corrected algorithm of wedges based on the lineattenuation was proposed. The lineattenuation coefficient of wedges was calculated firstly, then the actual path of radial beam in the wedge was calculated by the “linear fitting” according to the sizes of wedges section. Finally the wedge factors could be determined accurately, and the dose could be corrected. The data of the 45 degree wedge in the report of AAPM 55 were used to make the calculation and verificaton. Using this algorithm,amount of measured data were decreased, the accuracy of calculation could be assured, and the job of physician could be reduced,so this algorithm could be applied in the treatment planning system as an effective calculation method of wedges field.