高级检索
许丰, 郭建友, 於国兵, 徐平, 张建. 基于虚拟刻度原理对放射性体源效率刻度方法的发展[J]. 原子核物理评论, 2022, 39(3): 352-358. DOI: 10.11804/NuclPhysRev.39.2021087
引用本文: 许丰, 郭建友, 於国兵, 徐平, 张建. 基于虚拟刻度原理对放射性体源效率刻度方法的发展[J]. 原子核物理评论, 2022, 39(3): 352-358. DOI: 10.11804/NuclPhysRev.39.2021087
Feng XU, Jianyou GUO, Guobing YU, Ping XU, Jian Zhang. Development of Efficiency Calibration Method for Radioactive Volume Sources Based on Virtual Calibration Principle[J]. Nuclear Physics Review, 2022, 39(3): 352-358. DOI: 10.11804/NuclPhysRev.39.2021087
Citation: Feng XU, Jianyou GUO, Guobing YU, Ping XU, Jian Zhang. Development of Efficiency Calibration Method for Radioactive Volume Sources Based on Virtual Calibration Principle[J]. Nuclear Physics Review, 2022, 39(3): 352-358. DOI: 10.11804/NuclPhysRev.39.2021087

基于虚拟刻度原理对放射性体源效率刻度方法的发展

Development of Efficiency Calibration Method for Radioactive Volume Sources Based on Virtual Calibration Principle

  • 摘要: 采用蒙特卡罗模拟计算的方法,标定能够适用于不同HPGe探测器上的体源虚拟点源位置,需要对点源、体源的探测效率进行模拟计算。通过 241Am、137Cs、60Co点源和体源研究了HPGe晶体尺寸、类型对它们的虚拟点源位置的影响,模拟结果表明241Am虚拟点源位置随着探测器尺寸、类型的不同呈现明显的差异性,说明虚拟刻度原理对于探测器表面的小体积样品测量在低能区间是不可取的,最后采用 137Cs、60Co源得出体源高度与虚拟点源位置的半经验公式。通过实验室两台HPGe探测系统对尺寸Φ70 mm×65 mm标准土壤体源的探测效率进行计算,与模拟探测效率值、虚拟点源效率表征结果对比分析,验证了拟合一组表征虚拟点源位置的半经验公式是可行的。实验结果表明,对\gamma 能量区间在300~2 000 keV进行体源的虚拟刻度时,选择可靠的定标源即可建立体源和虚拟点源位置的关系。这为解决监测工作中样品与探测器之间重复进行探测效率校准的问题提供了新的途径。

     

    Abstract: Using the Monte Carlo simulation method, the virtual point source position of the volume source for different HPGe detectors can be quickly obtained. In the calculation, it is necessary to simulate the detection efficiency of the point source and the volume source. The effects of the size and type of HPGe crystals on their virtual point source positions were investigated through 241Am, 137Cs, 60Co point sources and volume sources. The simulation results show that the virtual point position of the 241Am source presents obvious differences. In the low energy range, the VPS calibration method is not valid in the measurement of small-volume samples of the surface of the detector. Finally, the semi-empirical formula for the height of the volume source and the position of the virtual point source is obtained by using the 137Cs and 60Co sources. In the test measurement, the detection efficiency of two types of HPGe detectors was calibrated using the standard soil source of the size of 70 mm×65 mm. Compared with the simulated detection efficiency curve and the calculation results of virtual point source efficiency, the validity of the semi-empirical formula was verified. In the gamma energy range of 300~2 000 keV, the relationship between the position of the volume source and the virtual point source can be established through a suitable radioactive source. This study provides a new approach in the radiation measurement to avoid the repeated detection efficiency calibration in detectors.

     

/

返回文章
返回