Abstract:
Since neutron sensitive microchannel plates (
nMCP) has high detection efficiency and the spatial resolution, combined with advanced readout electronics it can be a better choice for energy-resolved neutron imaging detectors. Compared with the matrix-doped
nMCP, the
nMCP based on Atomic Layer Deposition(ALD) has the advantages of less neutron sensitive material consumption and high secondary electron emission coefficient on the inner wall of the channel. Firstly, the typical neutron and gamma signal of
natGd-doped
nMCP were studied experimentally. Geant4 simulation and theoretical calculation were performed to optimize the pore diameter, wall thickness, bias angle and coating thickness of the coated
10B
2O
3 nMCP. It was shown that the thermal neutron detection efficiency was about 56% and the spatial resolution was about 22 μm when the coating thickness was 1 μm, the pore diameter was 10 μm, the wall thickness was 1 μm and the bias angle was 3°. The results are of great significance to the geometric parameter design of
nMCP used as energy-resolved neutron imaging detectors at CSNS.