一种用于RIBLL2的纵向场多次取样型电离室

唐述文, 段利敏, 孙志宇, 马朋, 鲁辰桂, 杨贺润, 张金霞, 胡正国, 徐瑚珊

唐述文, 段利敏, 孙志宇, 马朋, 鲁辰桂, 杨贺润, 张金霞, 胡正国, 徐瑚珊. 一种用于RIBLL2的纵向场多次取样型电离室[J]. 原子核物理评论, 2012, 29(1): 72-76. DOI: 10.11804/NuclPhysRev.29.01.072
引用本文: 唐述文, 段利敏, 孙志宇, 马朋, 鲁辰桂, 杨贺润, 张金霞, 胡正国, 徐瑚珊. 一种用于RIBLL2的纵向场多次取样型电离室[J]. 原子核物理评论, 2012, 29(1): 72-76. DOI: 10.11804/NuclPhysRev.29.01.072
TANG Shu-wen, DUAN Li-min, SUN Zhi-yu, MA Peng, LU Chen-gui, YANG He-run, ZHANG Jin-xia, HU Zheng-guo, XU Hu-shan. A Longitudinal Field Multiple Sampling Ionization Chamber for RIBLL2[J]. Nuclear Physics Review, 2012, 29(1): 72-76. DOI: 10.11804/NuclPhysRev.29.01.072
Citation: TANG Shu-wen, DUAN Li-min, SUN Zhi-yu, MA Peng, LU Chen-gui, YANG He-run, ZHANG Jin-xia, HU Zheng-guo, XU Hu-shan. A Longitudinal Field Multiple Sampling Ionization Chamber for RIBLL2[J]. Nuclear Physics Review, 2012, 29(1): 72-76. DOI: 10.11804/NuclPhysRev.29.01.072

一种用于RIBLL2的纵向场多次取样型电离室

A Longitudinal Field Multiple Sampling Ionization Chamber for RIBLL2

  • 摘要: 为兰州第二条放射性束流线(RIBLL2)研制了一台用于ΔE测量的纵向场多次取样型电离室。 利用3组分α源(239Pu为3.435 MeV, 241Am为3.913 MeV, 244Cm为4.356 MeV)对取样单元进行了测试, 确定了电离室的最佳工作电压为-500 V, 沉积能量为3.435 MeV时, 取样单元的能量分辨为271.4 keV(FWHM)。 利用Geant4对此电离室的整体性能进行了模拟, 表明可以对Z≥4的离子实现较好的粒子鉴别。 A longitudinal field MUltiple Sampling Ionization Chamber (MUSIC), which makes multiple measurements of energy loss for very high energy heavy ions at RIBLL2, has been constructed and tested with 3 constituent α source (239Pu: 3.435 MeV, 241Am: 3.913 MeV, 244Cm: 4.356 MeV). The voltage plateau curve has been plotted and -500 V is determined as a proper work voltage. The energy resolution is 271.4 keV FWHM for the sampling unit when 3.435 MeV energy deposited. A Geant4 Monte Carlo simulation is made and it indicates the detector can provide unique particle identification for ions Z≥4.
    Abstract: A longitudinal field MUltiple Sampling Ionization Chamber (MUSIC), which makes multiple measurements of energy loss for very high energy heavy ions at RIBLL2, has been constructed and tested with 3 constituent α source (239Pu: 3.435 MeV, 241Am: 3.913 MeV, 244Cm: 4.356 MeV). The voltage plateau curve has been plotted and -500 V is determined as a proper work voltage. The energy resolution is 271.4 keV FWHM for the sampling unit when 3.435 MeV energy deposited. A Geant4 Monte Carlo simulation is made and it indicates the detector can provide unique particle identification for ions Z≥4.