MICROMEGAS探测器初步测试

Preliminary Test of MICROMEGAS Gaseous Detector

  • 摘要: 介绍了一种利用鱼线做放大区间隔的MICROMEGAS探测器, 灵敏面积为50 mm ×50 mm。 在Ar+CO2(10%)气体条件下, 利用55Fe源5.9 keV的 X射线对MICROMEGAS探测器进行了初步测试: 计数率大于103 Hz的条件下, 计数率坪长达到280 V; 当栅极电压在800 V时能量分辨为30%; 当增益高于104时打火率低于10-4; PCB板读出条周期为400 μm条件下的位置分辨好于120 μm。 结果表明, 研制的MICROMEGAS探测器达到了初步设计的基本要求。 In this paper,A MICROMEGAS(MicroMeshGaseous Structure) detector with the amplification gap defined by stretching nylon fishing lines has been described . And its count plateau, gas gain and energy resolution in the condition of a 55Fe Xray source and ArCO2 (10%) mixture gases are measured as functions of mesh voltage. The count plateau approaches 280 V at the count rate of 103 Hz. The discharge rate is less than 10-4 when the gas gain is up to 104. The energy resolution is 30% (FWHM) based on -800 V mesh voltage. The position resolution is less than 120 μm at the condition of the sensitive area (50 mm×50 mm) Printed Circuit Boards (PCB) and its strip pitch of 400 μm. The results can satisfy the basic demand of MICROMEGAS detector preliminary design.

     

    Abstract: In this paper,A MICROMEGAS(MicroMeshGaseous Structure) detector with the amplification gap defined by stretching nylon fishing lines has been described . And its count plateau, gas gain and energy resolution in the condition of a 55Fe Xray source and ArCO2 (10%) mixture gases are measured as functions of mesh voltage. The count plateau approaches 280 V at the count rate of 103 Hz. The discharge rate is less than 10-4 when the gas gain is up to 104. The energy resolution is 30% (FWHM) based on -800 V mesh voltage. The position resolution is less than 120 μm at the condition of the sensitive area (50 mm×50 mm) Printed Circuit Boards (PCB) and its strip pitch of 400 μm. The results can satisfy the basic demand of MICROMEGAS detector preliminary design.

     

/