高级检索

基于混合Batho修正的RBM剂量计算方法在仿真头模实验中的剂量学验证

Verification of RBM Dose Engine in Experiment Using Emulational Head Phantom

  • 摘要: 基于混合Batho修正的规则束模型剂量计算方法可以根据放射治疗过程中加速器相关物理参数和从病人的医学影像数据中获取的人体组织非均匀信息计算出人体内的剂量场数据。在简要介绍该方法的基础上,通过将其计算结果与仿真头模的实测结果进行对比,初步证明了该方法可以作为一种较为可靠的快速剂量计算方法供临床治疗计划系统使用。The dose distribution can be calculated by the Regular Beam Model (RBM) dose engine based on the accelerator parameters and the inhomogeneity data of the patient for the radiotherapy planning. In this paper, the newly-developed mixed Batho correction based RBM is introduced, and then the calculation results are compared with the measured ones using the head phantom. The results demonstrate that the RBM dose engine can be used as a kind of reliable fast dose calculation tool in the clinical treatment planning system.

     

    Abstract: The dose distribution can be calculated by the Regular Beam Model (RBM) dose engine based on the accelerator parameters and the inhomogeneity data of the patient for the radiotherapy planning. In this paper, the newly-developed mixed Batho correction based RBM is introduced, and then the calculation results are compared with the measured ones using the head phantom. The results demonstrate that the RBM dose engine can be used as a kind of reliable fast dose calculation tool in the clinical treatment planning system.

     

/