大面积PIN探测器性能参数实验测量
Measurements of Parameters on Large Area PIN Detectors
-
摘要: 利用CFC-67型三通道γ源加速器测量了Ф50mm,Ф60mmPIN探测器的时间响应、线性电流及相对灵敏度,并与理论计算结果进行了比较,验证了实验结果的可靠性。Ф60 mm silicon PIN detector is a large area and high sensitive one which has been developed in near years. Using the type CГC-67 γ accelerator, the parameters of PIN detectors such as time response, linear current and relative sensitivity are measured. The experiment results are compared with the theory calculation. The PIN detector can reach high γefficiency of up to 10^-14-10^-16 C·cm2, the rise time is about 10 ns and the FWHM of the time pulse is about 35 ns. The γ efficiency of the PIN detector may provide the sensitive range between 10^-14-10^-16 C·cm^2 that scintillator + photo tubes detecting system and scintillator + photomul tiplier tubes detecting system is not developed.
Abstract: Ф60 mm silicon PIN detector is a large area and high sensitive one which has been developed in near years. Using the type CГC-67 γ accelerator, the parameters of PIN detectors such as time response, linear current and relative sensitivity are measured. The experiment results are compared with the theory calculation. The PIN detector can reach high γefficiency of up to 10^-14-10^-16 C·cm2, the rise time is about 10 ns and the FWHM of the time pulse is about 35 ns. The γ efficiency of the PIN detector may provide the sensitive range between 10^-14-10^-16 C·cm^2 that scintillator + photo tubes detecting system and scintillator + photomul tiplier tubes detecting system is not developed.