Calibration of LS301 Fastneutron Detector Efficiency
doi: 10.11804/NuclPhysRev.28.02.195
- Received Date: 1900-01-01
- Rev Recd Date: 1900-01-01
- Publish Date: 2011-06-20
-
Key words:
- 252Cf chamber /
- LS301 fast neutron detector /
- neutron detection efficiency /
- time of flight
Abstract:
Neutron detection efficiency of LS301 fastneutron detector was calibrated by measuring the neutron energy spectrum of 252Cf source, which has a standard fission neutron spectrum. A low mass, fast ionization chamber is used as the fission fragments detector in the timeofflight(TOF) spectrometer and afforded the start signal of neutron flight. The stop signal was offered by the anode of LS301. A measured TOF spectrum was turned to the neutron energy spectrum which will be compared with the standard one. Consequently, the fastneutron detection efficiency of LS301 was obtained. Calibration procedures of experimental and data processing was given. Relative detection efficiencies were obtained separately for threshold settings of 0.5, 0.7, 1.0 and 1.6 MeV for neutron energies under 10 MeV. Experimental results accorded with theoretical efficiency curves which were calculated with the Monte Carlo code NEFF50.
Citation: | ZHANG Wen-hui, LI Guang-wu, #, ZHANG Gang, CHEN Yu, YAN Guo-hong. Calibration of LS301 Fastneutron Detector Efficiency[J]. Nuclear Physics Review, 2011, 28(2): 195-199. doi: 10.11804/NuclPhysRev.28.02.195 |