Abstract: Based on the detail head structure designing provided by the manufacture, the ray transmission of VARIAN Clinic 2300C 6 MVX beam was simulated by Monte Carlo code, BEAMnrc. To accelerate the simulation efficiency, multiple optimized parameters were optimized. The simulated and the measured data were well coincident. The dose distribution of a complicated irregular field formed by MLC (MultiLeaf Collimators) was simulated, which are the base for further research work.
Citation: | LIN Hui, Wu Dong-sheng, LI Guo-li, JING Jia, YANG Zhu, . Simulating Varian 2300C 6 MV X Beam by Monte Carlo Code[J]. Nuclear Physics Review, 2008, 25(3): 254-259. doi: 10.11804/NuclPhysRev.25.03.254 |