Abstract: A Parallel Plate Avalanche Counter (PPAC) with weighted about 40 g was developed for isotope separator on-line Moessbauer Spectroscopy. It detects the internal conversion electrons emitted from the resonance Moessbauer nuclide. The signal-to-noise ratio of the spectra reaches 11 for 85% enriched ^119SnO2 cathode and absorber. It is an effective detector for on-line Moessbauer experiments. The perform ance of the counter was discussed in detail.
Citation: | HUANG Yan-jun, LIN Jun, XIA Yuan-fu. A Parallel Plate Avalanche Counter for Isotope Separator On-line Moessbauer Spectroscopy[J]. Nuclear Physics Review, 2006, 23(1): 66-69. doi: 10.11804/NuclPhysRev.23.01.066 |