高级检索
刘素萍, 龚建, 郝樊华, 伍怀龙, 胡广春, 向永春, 迮仁德, 章建华. 密封容器中钚材料同位素丰度和年龄的测量[J]. 原子核物理评论, 2005, 22(1): 127-130. DOI: 10.11804/NuclPhysRev.22.01.127
引用本文: 刘素萍, 龚建, 郝樊华, 伍怀龙, 胡广春, 向永春, 迮仁德, 章建华. 密封容器中钚材料同位素丰度和年龄的测量[J]. 原子核物理评论, 2005, 22(1): 127-130. DOI: 10.11804/NuclPhysRev.22.01.127
LIU Su-ping, GONG Jian, HAO Fan-hua, WU Huai-long, HU Guang-chun, XIANG Yong-chun, ZHE Ren-de, ZHANG Jian-hua. Measurement of Isotopic Abundance and the Age of the Plutonium Material in a Sealed Container[J]. Nuclear Physics Review, 2005, 22(1): 127-130. DOI: 10.11804/NuclPhysRev.22.01.127
Citation: LIU Su-ping, GONG Jian, HAO Fan-hua, WU Huai-long, HU Guang-chun, XIANG Yong-chun, ZHE Ren-de, ZHANG Jian-hua. Measurement of Isotopic Abundance and the Age of the Plutonium Material in a Sealed Container[J]. Nuclear Physics Review, 2005, 22(1): 127-130. DOI: 10.11804/NuclPhysRev.22.01.127

密封容器中钚材料同位素丰度和年龄的测量

Measurement of Isotopic Abundance and the Age of the Plutonium Material in a Sealed Container

  • 摘要: 根据相对效率法的原理,利用PC/FRAM分析软件,开展了用高分辨γ谱仪测量密封容器中钚材料的同位素丰度和年龄的实验研究.研究结果表明,同位素丰度和年龄的分析精度随能谱测量精度的提高而提高,如果能谱的测量精度不够,通过PC/FRAM分析得出的钚材料的同位素丰度和年龄有偏低的可能.因此,在能谱测量过程中,建议通过监测某个特征峰的峰面积计数来判定239Pu丰度和年龄的测量精度是否已满足测量要求.Based on the concept of relative detection efficiency, some experiments have been conducted to measure the isotopic abundance and the age of the plutonium material in a sealed container by a high-resolution γ-ray spectrometer. The PC/FRAM software analyzes the spectra. The study shows that the analysis precision of the isotopic abundance and the age will rise as the spectral measurement precision increases, and that If the spectral measurement precision is not enough, the analyzed results by the PC/FRAM software might be lower. Hence, the study suggests that a spectral peak is monitored to show lights on the spectral measurement precision during the process of spectrum acquirement.

     

    Abstract: Based on the concept of relative detection efficiency, some experiments have been conducted to measure the isotopic abundance and the age of the plutonium material in a sealed container by a high-resolution γ-ray spectrometer. The PC/FRAM software analyzes the spectra. The study shows that the analysis precision of the isotopic abundance and the age will rise as the spectral measurement precision increases, and that If the spectral measurement precision is not enough, the analyzed results by the PC/FRAM software might be lower. Hence, the study suggests that a spectral peak is monitored to show lights on the spectral measurement precision during the process of spectrum acquirement.

     

/

返回文章